33

7.HIGH FREQUENCY ELECTROTOME

Technical parameters

● The output power: 400W

● Six output modes: ① Cutting(400w )    ② blend cutl (200w)    ③ blend cut2(150w)

                                 ④soft Coag(120W)    ⑤ strong Coag(100W)    ⑥ bipolar Coag(70W)

● Atmospheric pressure range: 86.0-106.0kpa;

● Power supply:220V±22V, 50Hz±1;

● Working frequency: 360--160KHZ

● The rated power of the equipment: 880VA± 10%;

● Type of device: CF

● The machine power consumption: ≤1000VA

 

Functional features

● High definition digital display;

● The machine has automatic detection function, There is error code hint;

● It has a memory function, restart the emergence of last used power setting machine.

● It has a neutral electrode contact quality monitoring circuit, detection and patient contact plate area cabe effective contact area, evaluation of plates and skin system testing, once found the contact area isreduced to a dangerous level, automatically cut off the output system, and alarm.

● It can detect monopolar and bipolar neutral plates.

● It has automatic compensation function, stable output power, make the loss to a minimum.

● It can be used under water cutting, including lipid rich tissue cutting area.

● It can use with the endoscope and hysteroscopy.